Additional information
Brand | Hach |
---|---|
MPN | PD2P1 |
Model | General Purpose Insertion Mount pH Sensor |
Accuracy | ± 0.02 pH |
Body Material | PEEK |
Cable Connection | Analog |
Cable Length | 4.5 m PUR (polyurethane) 4-conductor with one shield, rated to 105 °C |
Calibration Method | Two point automatic, one point automatic, two point manual, one point manual. |
Communication | MODBUS |
Diameter | 34.9 mm |
Drift | 0.03 pH per 24 hours, non-cumulative |
Flow | max. 3 m/s |
Length | 196.3 mm |
Material | Ground Electrode: Titanium |
Mounting | Insertion |
Operating Temperature Range | -5 – 105 °C |
Pressure Range | max. 10.7 bar Sensor only (pressure range of mounting hardware to be checked separately) |
Range | -2.0 – 14.0 pH |
Repeatability | ± 0.05 pH |
Sensor Thread | 1" NPT (only at upper end) |
1" NPT (only at upper end) | pH-Wide Range |
Temperature Accuracy | ± 0.5 °C |
Temperature Compensation | automatic with NTC 300 Ω thermistor, or manually fixed at a user-entered temperature, |
Temperature Sensor | NTC 300 Ω thermistor for automatic temperature compensation and analyzer temperature readout |